22 GOSSEN METRAWATT GMBH
9.6 Alternating / Direct Voltage U
AC/DC
– Max. 253 V (feature F01)
Direct, alternating and pulsating voltages of up to 253 V can be measured
between the N (2) and L (3) connector jacks.
Select the U
AC/DC
measurement and start.
9.7 Resistance R (feature F01)
Resistance of up to 150 kΩ can be measured between the PE (1) and N
(2) connector jacks.
Select the R measurement and start.
9.8 Measurements with Accessories (feature F01)
9.8.1 Alternating Current I
Z
with Clip-On Transformer (feature F01)
Connection
Alternating current can be measured in two measuring ranges
(1 mA 10 A ~, 1 A 100 A~) with a clip-on current-voltage
transformer connected to the N (2) and L (3) jacks (e.g. the WZ12C).
Start the I
Z
measurement.
Select measuring range.